-3-
CAUTION
Do not touch the screen with your fingers. It is coated with a protective film.
c. Unplug connection (5).
d. Remove the eight screws (6). Carefully take off display panel.
e. Replace display panel in reverse order.
e.
Replace display panel in reverse sequence.
3.2 Remove/Replac e the Tare Kev Overlay.
a.
b.
c.
d.
Open the balance (see paragraph 3.1 a).
CAUTION
Do not touch the screen in the display with your fingers. It is coated with a
protective film.
Remove three clamps (1). Remove overlay from the display panel.
Take off the protective cover from the new overlay with touchpad.
Insert tare key overlay with the three ground connections in the display panel and fix it with glue. Place the
three clamps for the contact of the ground connection tare key overlay-display panel. Check new tare key
overlay. Use perenator to seal the joint around the tare key overlay and allow for several hours drying time.
4.
Recommended Spare Parts for Analytic Balance Model A 200 S.
Part Number
Description
Quantity
69 13019
Leveling Screw
1 ea
69 A20007-4
Overlay with Switch
1 ea
69 A200201
Front Glass Plate
1 ea
69 A20023-6
Rear Glass Plate
1 ea
6970920
AC/DC Adapter (115V) (US)
1 ea
69709170-0
LC-Display (A 200 S)
1 ea