4-11.
TO BLEED PRESSURE FROM TYPE I AND
TYPE 11 SERVICING UNITS
a.
Type I Servicing Unit - to bleed the air
pressure from a Type I servicing unit, proceed as follows
(See Figure 4-3)
(1)
Close charging valves (19) and close air
cylinder valves (2)
(2)
Place cam lever valves (5) in the up (open)
position
(3)
Open manifold shut-off valve (10) and
service valve (14).
(4)
Open bleed-off valve (18)
NOTE
If it is necessary to bleed-off the breathable air remaining
In the air cylinder (1), open air cylinder valves (2).
(5)
When all pressure has bled from the
breathable air system, close all open valves
b
TYPE I1 SERVICING UNIT (AIR/OXYGEN)
- To bleed the Air/Oxygen pressure from a Type II
Servicing Unit, proceed as follows (see Figure 4-4)
(1)
AIR half of Type 1I System - bleed AIR
pressure as per paragraph 4-11 a
(2)
Oxygen half of Type II system is as follows
(a)
Close charging valves (40) and close
oxygen cylinder valves (23)
(b)
Place cam lever valves (26) In the up
(open) position
(c)
Open manifold shut-off valve (31) and
service valve (35)
Open BLEED-OFF VALVE slowly Bleed
system slowly.
(d)
Open bleed-off valve (39)
NOTE
If It Is necessary to bleed-off the breathable oxygen
remaining in the oxygen cylinders (22), open oxygen
cylinder valves (23), lever cam valve (26) and bleed
valve (39)
(e)
When all pressure has bled from the
oxygen system, close all open valves
4-11.
GAS CYLINDER LOCATION ON TYPE I (AIR)
AND TYPE II1 (AIR/OXYGEN) SERVICING UNITS
NOTE
All component part descriptions as to location is from the
Panel View
a.
For the purposes of attaching flex hose
assemblies from the gas cylinders (Air or Air/Oxygen) to
the proper cam lever valve on the control panels, the
following number sequences shall apply (see Figure 4-5)
TYPE I (AIR)
(1)
Facing the control panel, from left to
right,
cam
lever
valves
will
be
numbered I through 8 Attach flex hose
line from numbered gas cylinder to the
correspondingly numbered cam lever
valve
TYPE II (AIR/OXYGEN)
(2)
AIR - See Type I above
(3)
OXYGEN - Facing the control panel
oxygen bulkhead is on the nght hand
side, numbered one (1) through four
(4)
Attach
flex
hose
line
from
numbered oxygen gas cylinders to
correspondingly numbered cam lever
valve
4-12
EMERGENCY BLEED PROCEDURE In case of
an emergency that would require immediate bleeding of
the breathable air servicing unit (Type I) or air/oxygen
servicing unit (Type II), open all cam lever valves (2.
Figure 4-1) and (2, 14, Figure 4-2) and open bleed
valves (14, Figure 4-1) and 6, 18, Figure 4-2)
4-9